Measuring skewness of red blood cell deformability distribution by laser ektacytometryстатья
Информация о цитировании статьи получена из
Web of Science,
Scopus
Статья опубликована в журнале из списка Web of Science и/или Scopus
Дата последнего поиска статьи во внешних источниках: 2 ноября 2014 г.
Аннотация: An algorithm is proposed for measuring the parameters of red blood cell deformability distribution based on laser diffrac- tometry of red blood cells in shear flow (ektacytometry). The algo- rithm is tested on specially prepared samples of rat blood. In these experiments we succeeded in measuring the mean deformability, deformability variance and skewness of red blood cell deformability distribution with errors of 10 %, 15 % and 35 %, respectively.